CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Carl Zeiss
Maintaining the device
ApoTome.2
28
423667-7144-001
05/2012
5.2
Maintaining the device
5.2.1
Performing checks
•
Make sure that the available line voltage corresponds to the line voltage values as labeled on the
ApoTome.2 control box.
•
Check power cable and power plug for damage.
•
In the event of perceptible damages, switch off the instrument and safeguard it against inadvertent
use. Have damages repaired by appropriately qualified staff.
5.2.2
Replacing fuses
Before replacing fuses, make sure to disconnect the power cable.
The fuse compartment is located at the rear panel
of the control box. It is combined with the
appliance connector and contains two fuses of the
type
T 2 A/H 250 V
.
•
Disconnect the power cable.
•
Pull out the fuse holder (5-1/
2
) forward. For
doing this, use a small screwdriver, if necessary.
•
Remove the defective fuses from the holder and
replace them by new fuses.
•
Slide the fuse holder back into the fuse
compartment (5-1/
1
) as far as it will go.
•
Connect the power cable.
Fig. 5-1
Replacing fuses
Содержание ApoTome.2