OPERATION
Carl Zeiss
Switching ApoTome.2 on/off
ApoTome.2
26
423667-7144-001
05/2012
4.2
Switching ApoTome.2 on/off
4.2.1
Switching on
•
Use the on/off switch (4-1/
3
) at the front panel of the ApoTome.2 control box to switch the device on.
•
The power indicator in the switch lights up.
Switch on the device approximately 15 minutes before the first image capture in order to allow
it to warm up to the necessary operating temperature.
The slider must be inserted in the microscope stand to warm up properly.
4.2.2
Switching off
•
After use, switch the device off by activating the on/off switch (4-1/
3
) at the front panel of the
ApoTome.2 control box.
•
The power indicator in the switch turns off.
4.3
Operating the slider
4.3.1
Inserting the slider
Carefully insert the ApoTome.2 slider in the plane of the luminous-field diaphragm of the reflected light
beam path of the microscope used. On insertion, a short beep signal is generated. In the first click-stop
position, conventional imaging is possible (free passage). In the second click-stop position, the
transmission grid is moved into the beam path.
4.3.2
Click-stop position 1
In click-stop position 1 of the ApoTome.2 slider, an open passage is in the reflected light beam path. In
this position the microscope can be operated in the conventional way. The slider does not affect any
function of the microscope stand used.
4.3.3
Click-stop position 2
In click-stop position 2, the transmission grid is positioned in the beam path of rays. All functions for the
adjustment and control of the instrument are operated via the software. For additional information, refer
to the Operating Manual of the software used.