5.4 Removing the airlock rod
When you do not wish to transfer specimens from the specimen chamber of the FESEM, you can
remove the airlock rod from the airlock.
1
Retract the airlock rod.
2
Change to
Vent
mode.
3
Open the airlock.
4
Remove the sample holder.
5
Remove the airlock rod:
a
Slightly turn the silver plug (
1
) counter-
clockwise
b
Remove the airlock rod.
6
Put the airlock rod onto the rack for storage.
7
Seal the orifice in the airlock window with
the blind plug.