6. Press
to save configuration and return to Run Screen.
7. Test General Alarm Relay output at→System→Service→Output Test -
page 144 or create a general alarm event to verify configured general
alarm relay energizes.
Note:
-
The General Alarm does not create an event log or control message regardless
if a General Alarm relay is enabled.
Name Devices/Sensors
Menus → System → Name Devices/Sensors
Names can be user configured for 5X00 instrument, sensor aux and E-Aux systems,
and relays. Default names are displayed at the local instrument when names are not
user configured. 5X00 default system names are: DO1-4 (depending on 5X00 model
number), Aux 1-6, E-Aux 1-8 and Relay 1-8. Default names for Aux 3-6 configured
as temperature systems are Temp 1-4. Instrument and Relay names are displayed at
many local instrument’s menus. System names are displayed at Run Screen, sensor
setup configuration, and control messages menus and screens. Because networked
instruments can use relays from remote networked instruments, it is important to
know that configured instrument and relay names are displayed differently at local
versus remote instruments’ relay menus. See additional information starting on
page 136.
Configure device and sensor names
1. Enter Name Devices/Sensors menu
Menus → System → Name Devices/Sensors>Name Devices/
Sensors Menu opens
Configuring the 5X00
2. Use
▲
▼
to scroll and highlight submenus.
3. Press
to select.
4. Use keypad
to configure name.
5. Press
to save configuration and return to Run Screen.
Notes:
-
See information regarding instrument and relay name conventions across a
node network - page 136.
-
Names can contain up to 20 characters.
-
Use AquaManager to easily configure names using the PC keyboard. See
AquaManager help.
-
Names uploaded in the temperature system name field within the temperature
system tab are displayed at the run screen, configuration menus, and control
messages menus.
-
Instrument, Aux & Sensor Names are not cleared at Reset to factory menus.
-
Relays names are cleared at “reset relays”, “reset all systems” and “reset to fac-
tory/reset all” service menus.
-
5X00 default system name is displayed at event, sensor, and calibration logs,
regardless if system name is user configured.
-
At alpha/numeric keypad enter charcter then press “Clear” at keypad and then
press
to clear user configured names.
-
Aux and E-Aux system names can also be configured at: Menus → System →
Configure Aux /Configure E-Aux→Aux 1-6/Eaux 1-8→Name
-
Because of font limitations, 1X2 and 2X3 display formats do not show all 20
characters of a 20 character name. Typically only 13 characters are displayed.
-
The instrument name is not shown at Run Screen. The instrument name is
shown at relay enable menus for a local instrument only.
-
Certain menus (i.e. configure aux) display up to 18 characters of the device
name. “...” is shown at the menu if the name exceeds 18 characters.
Configuring the 5X00