YSI 5X00
123
YSI 5X00
122
Date/Time
Menus → System → Date/Time
It is important to configure the correct date and time in the 5X00. Four 10-event
daily timers and four feed timer run based on the 5X00 date and time. Logs record
events based on the 5X00 clock. Date and time display formats are user defined.
1. Enter Date/Time menu
Menus → System → Date/Time
2. Verify 5X00 current date and time.
3. If necessary, use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Configure submenus.
6. Press
to save setting and return to Run Screen or press
to System menu.
Notes:
-
Changes to Date and Time menus take place immediately and are saved prior
to
.
-
Configured date and time formats are carried throughout menus and logs.
-
Any active (with energized relay) timer system, including 10-event daily
timers, feed timers, and/or dosing timers, are NOT recalculated when the
5X00 time and/or date is changed. Once a timer relay has been energized, it
runs for the duration and does not reset based on changes to the 5X00 date
and time. Example - timer 1, event 1 is configured to come on at 9:00am for
10 minutes. Timer relay energizes at 9:00am. At 9:05am the 5X00 time is
changed to 10:05am (for daylight savings adjustment). The timer relay will
de-energize after five additional minutes. The 5X00 time will be 10:10am. The
event log will reflect the configured 5X00 time when the event occurred. In
this example, 9:00am for the start and 10:10am for the end.
-
The time must be manually changed for daylight savings time adjustments.
Daylight savings note: Sensor data log records are not overwritten when the
Configuring the 5X00
clock is changed unless the instrument is power cycled after changing the
data and time. Example - the data log interval is set to 15 minutes. Current
instrument time is 6:00 am. The clock is adjusted to 5:00 am for daylight
savings time. The 5X00 sensor log contains four 15 minute interval records
from 5:00 am to 6:00 am from prior to time change. Sensor data logs are not
overwritten with new values for the repeat of 5:00 am to 6:00 am 15 minute
intervals unless the instrument is power cycled.
-
Menu configurations are saved to the system by pressing
to return to
the Run Screen and not when the change is made in the menu. Exception -
Date and time and clear data logs are updated immediately to the system, not
when
is pressed.
Display Settings
Menus → System → Display Settings
Sensor and Aux conditions are displayed at the Run Screen. Use the Display Set-
tings menu to configure how many systems display on one page at the Run Screen.
Up to six systems can be displayed on one Run Screen page. Multiple Run Screen
pages scroll at user configured speed (1-65535 seconds).
Configuring the 5X00