4. Ensure that the rubber gasket is seated between front panel and enclosure
whenever enclosure is closed.
5. Replace front panel making sure that ribbon cable is not trapped in the
gasket channel before inserting mounting screws.
6. Tighten mounting screws, making sure not to cross thread. The screws
are stainless steel and the receiving threads are brass. Do not over-
tighten!
WARNING: The 5X00 utilizes sensitive solid-state devices that can be damaged
by static shock. Installers must observe accepted ESD (Electro-Static Discharge)
procedures while connecting cabling to the 5X00 I/O plate or any other internal
component or damage may result.
AVERTISSEMENT : le modèle 5X00 utilise des appareils transistorisés sensibles
pouvant être endommagés par des décharges d’électricité statique. Les installateurs
doivent observer les procédures de protection acceptées contre les décharges
électrostatiques lors du câblage à la carte d’E/S du modèle 5X00 ou à tout autre
composant interne afin d’éviter tout risque de dommage.