Service
Menus → System → Service
The Service menu includes submenus to configure and Reset Clean Probe timer,
Reset Systems to factory defaults, and test outputs including relays and IOEM
4/20mA outputs.
Reset Clean Probes
Menus → System → Service→ Reset Clean Probes
Enable clean probe timer, configure the duration, and reset clean probe days remain-
ing from Reset Clean Probe menu. Selecting reset at this menu resets clean probe
timer to the user configured value.
icon displays at Run Screen when the
clean probe timer expires.
Configuring the 5X00
Reset Clean Probe timer at the
menu or at the Reset Clean Probes menu.
When prompted “Are you sure?”, select “Yes” to reset the clean probe timer to the
user configured number of days.
Reset To Factory
Menus → System → Service→ Reset To Factory
The Reset to Factory menu allows resetting of some or all of 5X00 systems to factory
defaults. Example - if Clean Probe Timer is user configured to 60 days, performing a
factory reset will reset the Clean Probe Timer to 90 days which is the factory default
setting and not the user configured value. See
Appendix B Menu Maps
starting on
page 225 for factory default settings for all systems.
Notes:
-
“Reset All” and “Reset All Systems” resets time format to 12 hour and date
format to MMDDYYYY.
-
It is strongly recommended that user configuration be maintained in a log
by the user.
-
Instrument, aux, sensor, and suffix names are not reset at any reset to factory
menus. Relay names are reset at “reset relays” and “reset all” menus.
-
Control icons and control menu messages are displayed for 30 seconds after
pressing
from reset to factory menus for any system that was active
(energized relay) prior to factory reset.