Control Menu
Press
at the Run Screen to display a list of energized sensor and aux 5X00
system relays. The Control menu identifies the active sensor or aux system and
control and/or alarm condition. See Control mode and alarms - page 152 for
additional information.
Notes:
-
Press
to display time stamped detail of control and alarm relay(s).
-
Alarm message appears at the Control menu for aux and sensor systems when
alarms are enabled regardless if an alarm relay is assigned.
-
Active timer (10 event timer and feed timer) systems are not displayed at
control menu.
-
General alarm systems are not displayed at the Control menu regardless if
general alarm relay is enabled.
-
Use
▲
▼
to scroll list when scroll bar appears at window.
When no 5X00 control or alarm system(s) are active, the below screen is displayed.
Msg’s Menu
Calibration, Sensor, and Event logs contain 5X00 system information. The 5X00
Sensor log stores 2000 records. The Calibration log stores 500 records. The Event log
stores 1250 records. Once the log memory is full, the oldest records are overwrit-
ten with more recent records. The number shown in ( ) at the Msg’s menu reflects
the total number of records in the log. The 25 most recent records for each log are
displayed at the
menu. Use
▲
▼
to scroll and press
to view time
stamped detail record.
Use AquaManager to capture all records from 5X00 logs and store them to a PC
database. Polling times should be determined by logging intervals. Example - if the
sensor log is recording values every minute and the sensor log holds 2000 records,
polling should occur at least every 33.3 hours. This allows for all records to be
saved to the PC before being overwritten on the 5X00. Set the data log interval so
that analyzing data based on the configured interval will be helpful. Poll at intervals
that don’t result in gaps in data.
A Reset Clean Probe message will appear at the Msg’s Menu if the clean probe timer
expires. Selecting “Yes” at “Are you sure?” resets the clean probe timer to the user
configured number of days.
Note:
-
User configured sensor and relay names do not appear in log menus.