aquaMAX
TM
– Ultra 370 Series
- 23 -
2) UV lamp replacement
①
Press
Service
key on the standby mode
②
Press
Service
key 2 times first and
Product
key to run Pressure Down. (Make sure to have O.K sign)
③
Perform the process above 2~3 times to remove inside pressure.
④
Turn off the system. Loosen a captive screw on the up-side of a UV lamp support and pull forward as
following Fig.4-2-(a)
⑤
Detach the power connector as following Fig. 4-2-(b).
⑥
Take off the tightener to separate a connecter cap connected to a lamp power code on the upside of UV
chamber. See the Fig. 4-2-(c)
⑦
Carefully lift up the cap in straight with UV chamber as following Fig. 4-2-(d)
⑧
Grab the connector cap with one hand and carefully detach the UV lamp with the other hand.
⑨
Get a new UV lamp and insert it to the connect inside of cap. Try not to leave any finger prints or
contaminants on the UV lamp body.
⑩
There is a glass tube inside of UV chamber, and keep it our of produced water. If there is water in it, dry
it out.
⑪
Make sure a UV lamp firmly connected, and insert the UV chamber back, and follow the above
⑤
~
⑦
in reverse. Carefully install all back with no produced water leakage .
(a) UV lamp support (b) Detachment from power connector