aquaMAX
TM
– Ultra 370 Series
- 21 -
Chapter 4. Maintenance
1. The aquaMAX
TM
-Ultra maintenance
2. Filter safekeeping
3. Filter replacement
4. Temperature resistivity data acquisition by RS232C
5. Service
- Resistivity cell calibration at the filter replacement
4-1. The aquaMAX
TM
-Ultra maintenance
- To keep the water quality of feed water
: Feedwater for Ultrapure water system should be R.O(Reverse Osmosis) grade water or D.I(Deionized) water.
Therefore, to use filter longer, the feedwater needs to be fresh and kept in good condition.
- Periodic operation
: If the system keeps power on, it can increase the time of use because recirculation runs every 1 hour.
4-2. Filter safekeeping
For Ultra Pack or UF Filter : If the system is not used for a long time, keep the water inside of filter and close
the input and output with the cap. Keep the air out of filters. When it comes to using again, run “ stabilization”
on service menu a few times to remove air or contaminants. Keep the UV lamp in the safe place.
4-3 Filter replacement
1) ULTRA-Pack replacement
①
Press
Service
key on the standby mode
②
Press
Service
key 2 times first and
Product
key to run Pressure Down.
③
Press
Service
key 6 times, and Product key 2 times to perform CHANGE FILTER.
④
After it displays OK, it turns to Standby mode, then, open the front door.