Hand-press down the clip connecting the installed filter and fitting as following Fig 4-1-(b). Once
hearing the "click" sound from the clip, grab the filter and slowly pull it out.
(a) Ultra Pack Installed (b) Ultra Pack removal (c) After the filter removal
[
Fig. 4-1
]
Ultra Pack Replacement
⑥
After the filter removal, hand-press down the clip as following Fig 4-1-(c) by making a “click” sound to
install new one. There can be small amount of water leaking during the filter replacement.
⑦
Remove the cap ( The cap for input and output of filters) ,then, install Ultra Pack 1 and 2 each by no.
on the instrument. Insert it to fix into the fitting until you hear “click” sound.
⑧
Slightly pull out the installed filter to make sure it’s fixed.
⑨
Press
Operate
button to run "System Check"(Flushing Ultra Pack). The flushing runs for 20 min.
If the resistivity is still lower than 13M
Ω·
cm, run it again.
⑩
The system checks the resistivity on recirculation after System Check is done.
⑪
Once the product data gets to 13M
Ω
at operate mode, you may press
Production/Recirculation
button to produce water.