aquaMAX
TM
– Ultra 370 Series
- 11 -
4. Power Supply
Connect to 110/220V code depending on supplied power.
Both 110V and 220V power is available with free voltage support.
5. Operating Test
After Switch On, run “Self test” and check the display of the name of model and logo.
Press
Operate
key and check it turns to Ready mode.
6. Connection of feed water and reject water
Feed Water Connection Reject Water Connection Connection Completed
[
Fig. 2-4
]
Tube connections for Feed Water and Reject Water
Use for R.O(Reverse Osmosis) grade water or D.I(Deionized) water as feed water.
Inside of reservoir tank should be organic-free. Put the reservoir tank higher place than the instrument.
After connecting the tubing (3/8” Teflon tubing) connect to feed water reservoir, fit into FEED inlet on the rear
side. For the connection of Reject water, use 1/4” Teflon tubing. The other end of tubing needs to be placed
where for drain or for wastewater.