13. If the state shows “
reject
” value(2), the entire sanitization procedure is not complete.
-
Push the
Operate
key to enter “
feed water
” state, and then push the
Operate
key to enter to try again
the “
first sanitize
”.
14. The sanitizing procedure must be continued until the state has “
stable
” value
15. If the state shows “
stable
” value(1), the entire sanitization procedure is now complete.
-
Push the
Operate
key to enter “ready”.
(a) (b)
Figure. Sanitization procedure for UF filter
1) Open the cap for UF sanitization port as fig (a)
2) Insert NaOH into UF sanitization port