aquaMAX
TM
– Ultra 370 Series
- 14 -
Chapter 3. Operation
3-1.Key Operations
[
Display
]
- The resistivity of produced water is shown. Press
Display
key during operation to see the temperature.
Press once more to get back the resistivity display.
[
Service
]
- This key has various functions up to 6 (or 7) on
「
Standby Mode
」
. Press
Service
key to move to each
function and press
Production/Recirculation
key to run.
1. Stabilization
It is to remove air in filters and tubes at the first installation or replacement. The system produces the
produced water for 1 min. ( The REJECT valve is opened for 10 sec, and waste water will be drained through
the REJECT line at the back-side.
■
S E R V I C E
■
S t a b i l i z a t i o n . .
■
S E R V I C E
■
C o m p l e t e d. .
2. Press Down
It’s used to reduce inlet pressure when any parts of filters need to be replaced.( Waste water will be drained
through the REJECT line at the back-side.
■
S E R V I C E
■
P R E S S D O W N