Error Message
-HIGH PRESSURE ERROR
If there is high pressure occurred in the system during OPERATE/PRODUCT, decompressing gets started
with this message. If the high pressure still remains even though the decompressing, please check if there is
anything wrong with the system and filters.
■
HIGH PRESSURE ERROR
■
S Y S T E M C H E C K
- Over used Filter Error
In case of overuse of filter life, the system stops the operation with FILTER time CHECK message when the
power on and gets backs to
「
Standby
」
mode. Normal operation is still available, but To replace filters, run
CHANGE FILTER function on Service menu. “CHANGE FILTER” message keeps popping up every 30 min.
■
O p e r a t e
■
■
STANDBY
■
U L T R A P A C K C H A N G E
F I L T E R T I M E C H E C K
Reset the filter time.
■
S E R V I C E
■
C H A N G E F I L T E R
- If the resistivity is lower than 13M
Ω·
cm, the production is not possible and the system displays the message
as below.
■
O p e r a t e
■
N o t S t a b l e ! !
If the resistivity drops down under 10 M
Ω·
cm during operation , the system stops producing to re-circulate. If
the resistivity is higher than 13M
Ω·
cm, production is available any time by pressing
PRODUCTION
key.
■
O p e r a t e
■
N o t S t a b l e ! !