<12. Troubleshooting>
12-9
Alarm 10: Cold Junction Temperature Alarm
This alarm is generated when the temperature of the cold junction located at the terminal block of
the detector falls below -25°C or exceeds 155°C. Check the following:
Display “C.J.temperature” in the Detailed-data display. If “C.J.temperature” is indicated as 200°C
or -50°C, the following can be considered.
(1) Breakage of the cold junction signal wires between the converter and the detector occurred,
or the cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the cold junction signal wiring are shorted out in the
wiring extension or at the connection terminals.
(3) A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
(4) A failure of the electrical circuits inside the converter occurred.
If “C.J.temperature” exceeds 150°C or falls below -20°C, the following can be considered.
(1) The temperature of the detector terminal block is out of the operating temperature range
(-20 to 150°C).
(2) A failure of the cold junction temperature sensor located at the detector terminal block
occurred.
(3) A failure of the electrical circuits inside the converter occurred.
<Locating cause of alarm and countermeasures>
Before proceeding to the following troubleshooting procedure, examine whether or not the
temperature of the detector terminal block is out of the operating temperature range.
The operating temperature range varies with the type of detector. If the detector terminal block
is out of its operating temperature range, take the measure to lower the temperature, such as
situating it so that it is not subjected to radiant heat.
The case where the Model ZR22 Detector is used:
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 5 and 6 of the detector and measure the resistance
between these terminals. If the resistance value is out of the range of 1 to 1.6 kΩ, the cold
junction temperature sensor is considered to be faulty. Replace that temperature sensor
with a new one.
(3) If the resistance value is within the above range, the cold junction temperature sensor
seems to be normal. Check whether or not the cable is broken or shorted out, and whether
the cable is securely connected to the terminals. Also, check that the resistance of the wiring
between the converter and detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa Electric Corporation.
IM 11M13A01-02E
Содержание ZR22S
Страница 11: ...Blank Page...
Страница 19: ...Blank Page...
Страница 23: ...Blank Page...
Страница 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Страница 70: ...Blank Page...
Страница 134: ...Blank Page...
Страница 164: ...Blank Page...
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Blank Page...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......