<11. Inspection and Maintenance>
11-5
11.1.3 Replacement of the Heater Unit (Assembly)
This Subsection describes the replacement procedure for the heater unit.
The sensor or ceramic heater-furnace core internal structure is subject to fracturing, so do NOT
subject it to strong vibrations or shock. Additionally, the heater unit reaches high temperatures
and is subjected to high voltages. So, maintenance services should be performed after the power
is off and the heater unit temperature has returned to normal room temperature.
For details, refer to IM11M12A01-21E “ Heater Assembly ”.
NOTE
If the heater strut assembly cannot be removed because a screw has fused to its thread, one of
our service representatives can fix it.