<10. Other Functions>
10-10
10.4.5 Default Setting
When the analyzer is delivered, or if data are initialized, the blow back settings are by default, as
shown in Table 10.4.
Table 10.4 Blow back Default Setting
Item
Default setting
Mode
No function (“invalid”)
Hold time
10 minutes, 00 seconds
Blow back time
10 minutes, 00 seconds
Interval
30 days, 00 hours
Start date
01/ 01/00
Start time
0:00
10.5 Parameter Initialization
Parameter settings can be initialized to the factory default settings. Initialization can be done
for all parameters or for individual parameters. The parameters that can be initialized and their
defaults are listed in Table 10.5.
To initialize the set data, follow these steps:
(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Commissioning”, next “Others” then “Defaults”. A display like Figure 10.11. appears.
(3) Select the desired item to be initialized then a display like Figure 10.12 appears.
Select “Defaults start” then initialization starts.
Display
Calibration
Blow back
mA-outputs
Alarms
Contacts
Others
All data
Enter
Defaults
F10-11E.ai
Enter
Defaults start
Cancel
Defaults
Display
F10-12E.ai
Figure 10.11 Defaults Display
Figure 10.12 Initialization Start Display
CAUTION
• Do not turn off the power during initialization. Otherwise, initialization will not be performed
properly.
IM 11M13A01-02E
Содержание ZR22S
Страница 11: ...Blank Page...
Страница 19: ...Blank Page...
Страница 23: ...Blank Page...
Страница 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Страница 70: ...Blank Page...
Страница 134: ...Blank Page...
Страница 164: ...Blank Page...
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Blank Page...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......