<7. Startup>
7-10
(2) Span gas concentration
With “Span gas conc” selected in the Calibration setup display, display the Numeric - data
Entry display and enter an oxygen concentration value for the span gas calibration; If
instrument air is used, enter 02100 for a 21 vol%O
2
value.
When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas),
use a portable oxygen analyzer to measure the actual oxygen concentration, and then enter
it.
CAUTION
If instrument air is used for the span gas, dehumidify the air by cooling to a dew point of - 20°C
and remove any oil mist or dust.
7.12.2 Manual Calibration
n
Preliminary
Before performing manual calibration, be sure that the ZA8F Flow Setting Unit zero gas flow
setting valve is fully closed. Open the zero gas cylinder pressure reducing valve so that the
secondary pressure equals sample gas plus approx. 50 kPa (or sample gas pressure plus
approx. 150 kPa when a check valve is used, maximum pressure rating is 300 kPa).
This applies even if you are using the ZR40H Automatic Calibration Unit.
n
Calibration Procedures
This manual assumes that the instrument air is the same as the reference gas used for the span
gas. Follow the steps below to conduct manual calibration:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Calibration” in the Execution/Setup display. In doing so, the Calibration display
as in Figure 7.21 appears.
Enter
Calibration
Manual calibration
F07-21E.ai
Check the span gas
concentration value
Next
Cancel calibration
Change value: 21.00%
Enter
Manual calibration
F07-22E.ai
Figure 7.21 “Calibration” Display
Figure 7.22 “Manual Calibration” Display
(2) Press the [Enter] key to select “Span calibration”. The Manual calibration display shown
in Figure 7.22 then appears. Check that the oxygen concentration for the span gas in this
display coincides with the oxygen concentration in the calibration gas actually used. If the
check results are assumed to be OK, select “Next” in the Manual calibration display.
IM 11M13A01-02E
Содержание ZR22S
Страница 11: ...Blank Page...
Страница 19: ...Blank Page...
Страница 23: ...Blank Page...
Страница 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Страница 70: ...Blank Page...
Страница 134: ...Blank Page...
Страница 164: ...Blank Page...
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Blank Page...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......