<11. Inspection and Maintenance>
11-2
IM 11M13A01-02E
11.1 Inspection and Maintenance of the Detector
11.1.1 Cleaning the Calibration Gas Tube
The calibration gas, supplied through the calibration gas inlet of the terminal box into the detector,
flows through the tube and comes out at the tip of the probe. The tube might become clogged
with dust from the sample gas. If you become aware of clogging, such as when a higher pressure
is required to achieve a specified flow rate (600 ± 60 ml/min), clean the calibration gas tube.
To clean the tube, follow these steps:
(1) Remove the detector from the installation assembly. Remove the Flame Arrestor Assembly
from Sensor Assembly (See Figure 11.2)
(2) Following Subsection 11.1.2, later in this manual, remove the four bolts (and associated
spring washers) that tighten the sensor assembly, and the pipe support as well as the
U-shaped pipe.
(3) Use a rod 2 to 2.5 mm in diameter to clean the calibration gas tube inside the probe.
In doing this, keep air flowing from the calibration gas inlet at about 600 ml/min and insert
the rod into the tube (3-mm inside diameter). However, be careful not to insert the rod
deeper than 40 cm for a general-purpose detector, or 15 cm for a high temperature detector.
(4) Clean the U-shaped pipe. The pipe can be rinsed with water. However, it should be dried out
thoroughly before reassembly.
(5) Restore all components you removed for cleaning. Follow Subsection 11.1.2 to restore all
components in their original positions. Be sure to replace the O-ring(s) with new ones.
F11-1E.ai
Calibration gas tube
Exploded view of components
Rod
(with outside diameter
of 2 to 2.5 mm)
Figure 11.1 Cleaning the Calibration Gas Tube
11.1.2 Replacing the Sensor Assembly
The performance of the sensor (cell) deteriorates as its surface becomes soiled during operation.
Therefore, you have to replace the sensor when its life expectancy expires, for example, when
it can no longer satisfy a zero correction ratio of 100±30% or a span correction ratio of 0±18%.
In addition, the sensor assembly is to be replaced if it becomes damaged and can no longer
operate during measurement.
If the sensor becomes no longer operable (for example, due to breakage), investigate the cause
and remedy the problem as much as possible to prevent recurrence.
Содержание ZR22S
Страница 11: ...Blank Page...
Страница 19: ...Blank Page...
Страница 23: ...Blank Page...
Страница 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Страница 70: ...Blank Page...
Страница 134: ...Blank Page...
Страница 164: ...Blank Page...
Страница 190: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Blank Page...
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......