9-5
IM 253421-01E
Using External Input/Output
9
9.3 D/A Output (optional)
Setting the Integration Preset Time
The D/A output of integrated values will be 5.0 V FS when the rated range has been
input consequently during the preset integration time (rated integration time).
Setting range: 0.00.00 (0 hrs 0 min 0 sec) to 10000.00.00 (10000 hrs 0 min 0 sec)
The initial value is 1.00. When 0.00.00 is set, the D/A output value will be 0 V.
Note
•
The displayed values of V (voltage), A (current), W (active power), VA (apparent power), var
(reactive power), Vpk (voltage peak), and Apk (current peak), while the MAX hold function
(see section 4.8) is enabled, will be the maximum values (MAX) that are held. The values for
D/A output are also set to the maximum values (MAX) that are held.
•
The D/A output of each output item is adjusted so that 5.0 V is output when the value
corresponding to the range rating of the voltage, current, or power is applied.
•
Even when scaling constants for voltage, current, and power are specified, the D/A output is
adjusted so that 5.0 V is output when the value corresponding to the range rating is applied.
•
When
(computation) is specified, the D/A output is 0 V.