9-1
IM 253421-01E
Using External Input/Output
9
9.1
Remote Control and D/A Output Connector
(optional)
Using the remote control and the D/A output connector, this instrument can be remotely
controlled and D/A output can be done. The connector’s pin sequence and signal
assignment is as follows.
Connector’s Pin Sequence
24
13
1
12
(Rear panel)
Pin Assignment
remote control, 4 channel D/A output
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
DIGITAL COM
EXT HOLD
EXT START
EXT RESET
No Connection
DA 3ch
DA 1ch
DA COM
DIGITAL COM
EXT TRIG
EXT STOP
INTEG BUSY
DA 4ch
DA 2ch
DA COM
Pin No.
Signal
Pin No.
Signal
(Output)
(Output)
(Output)
(Output)
(Input)
(Input)
(Input)
(Input)
(Input)
(Output)
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
No Connection
0.01
µ
F
+5 V
L
H
+5 V
10 k
Ω
100
Ω
100
Ω
: 0 to 0.8 V
: 2.0 to 5 V
L
H
: 0 to 0.4 V(8 mA)
: 2.4 to 5 V(–400
µ
A)
Remote control: input circuit
TTL level
TTL level
Remote control: output circuit
WARNING
The connectors used in this function have protective covers. When the covers
are removed or when using connectors, the voltage ratings across the
measuring input and the ground become as follows:
Voltage across A,
±
(V and A side) input terminals and ground 400 Vrms max.
Voltage across V terminal and ground 600 Vrms max.
Put the protective cover on the connector when this function is not used.
Chapter 9 Using External Input/Output