14-14
IM 253421-01E
AOUTput:IRTime
Function
Sets the preset integration time for D/A output
of integrated values, or queries the current
setting.
Syntax
AOUTput:IRTime {<NRf>,<NRf>,<NRf>|
<String>}
{<NRf>,<NRf>,<NRf>}=0,0,0 to 10000,0,0
{<String>}=HHHHH:MM:SS
HHHHH hours
MM minutes
SS seconds
Example
AOUTPUT:IRTIME 1,0,0
AOUTPUT:IRTIME "1:00:00"
AOUTPUT:IRTIME?
→
:AOUTPUT:IRTIME 1,0,0
AOUTput:PRESet
Function
Initializes the output items for D/A output.
Syntax
AOUTput:PREset {NORMal|INTEGrate}
NORMal=default for normal measurement
INTEGrate=default for integration
Example
AOUTPUT:PRESET NORMAL
Description
See section 9.3 for a description of default D/A
output items for normal measurement and
integration.
AOUTput?
Function
Queries all the settings relating to D/A output.
Syntax
AOUTput?
Example
AOUTPUT?
→
:AOUTPUT:CHANNEL1 V,1;
CHANNEL2 A,1;CHANNEL3 W,1;
CHANNEL4 VHZ,1;IRTIME 1,0,0
AOUTput:CHANnel<x>
Function
Sets the D/A output item, or queries the current
setting.
Syntax
AOUTput:CHANnel<x> {<normal measurement
function>,(<NRf>|ELEMent<1>)|OFF}
<x>=1 to 4 (for /DA4)
<normal measurement function>={V|A|W|VA|
VAR|PF|DEGRee|VHZ|AHZ|WH|WHP|
WHM|AH|AHP|AHM|MATH|VPK|APK}
Example
AOUTPUT:CHANNEL1 V,1
AOUTPUT:CHANNEL1?
→
:AOUTPUT:CHANNEL1V,1
AOUTPUT:CHANNEL2?
→
:AOUTPUT:CHANNEL2 OFF
14.3 Commands
14.3.2 AOUTput Group
The commands in the AOUTput group are used to make settings relating to, and inquires about D/A output. This
allows you to make the same settings and inquiries as can be set using the lower menus of [OUTPUT]-“dA” or
[INTEG SET]-“dAtimE”. The commands in this group are available only if the /D4 option is installed.
:AOUTput
:
PRESet
<Space>
NORMal
INTEGrate
CHANnel
<x>
<Space>
<Normal meas. function>
,
<NRf>
ELEMent
<x>
OFF
?
IRTime
<Space>
<NRf>
,
<NRf>
?
?
<Character string>
;
,
<NRf>