4.12 Selecting the Number of Displayed Digits
Keys
DC
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
M W
h
k
A
h
m V
Hz
M W
k
A
deg
var
TIME
FUNCTION
AUTO
MODE
TRIG
MAX HOLD
HOLD
CAL
INTEGRATOR
HARMONICS
INTERFACE
OUTPUT
SHIFT
REMOTE
MEMORY
INTEG SET
ENTER
A RANGE
V RANGE
START
LOCAL
SETUP
STOP
RESET
AUTO
FUNCTION
FUNCTION
m V
PF
M W
k
A
m V
VA
V MEAN
RMS
A OVER
V OVER
SAMPLE
A
B
C
MODE
Displays
relevant keys
Procedure
•
Operate the instrument by following the thick lines in the menu below.
•
Press the ENTER key to confirm a selection or setting.
•
To leave the current menu in the middle of the operation, press the key indicated in step 1.
The confirmed settings up to that point are kept.
ENTER
3.
SETUP
1.
ENTER
4.
2.
5
.
Select the number of displayed digits
(Display C)
(Display C)
End
Explanation
Function used to select the number of displayed digits
You can select the maximum number of displayed digits for values such as V
(voltage), A (current), W (active power), VA (apparent power), var (reactive power),
Vpk (voltage peak), Apk (current peak), PF (power factor), VHz (voltage frequency),
AHz (current frequency), four arithmetical operation values, crest factor, and harmonic
analysis values (voltage, current, and active power). The initial value is Lo.
• Hi
The number of displayed digits is set to 5 (99999).
• Lo
The number of displayed digits is set to 4 (9999).
Note
•
The actual number of displayed digits may be smaller than the maximum number of displayed
digits depending on the combination of the voltage range and current range and the
automatic digit carrying operation.
•
Some items such as the phase angle, integrated value, integration time, and relative
harmonic content are not affected by the maximum number of displayed digits specified in
this section. For details, see the sections describing each item.