10-7
IM 253421-01E
Using the GP-IB Interface (Optional)
10
Output Format when Self Selected
Up to 14 normal measured/computed data can be output simultaneously, and the user
is allowed to choose any output information type for those 14 data. Each output block
is of the following format.
ch.1
,
ch.2
ch.3
ch.4
Terminator
,
,
ch.5
,
ch.6
ch.7
ch.8
Terminator
,
,
ch.9
,
ch.10
ch.11
ch.12
Terminator
,
,
ch.13
,
ch.14
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
Each output block usually consists of five lines (six in case of recall) including the
block end line “END”. However, if all output types on a line are set to “no output”, this
line will be omitted, reducing the number of output lines by one. For example, if all
output items of ch.9 to ch.12 are set to “no output”, line 4 in the above example will be
omitted.
Furthermore, if any channel on a line is set to “no output”, all data following this
channel on the line will be shifted forward. For example, if the ch.2 on line 1 is set to
“no output”, data of ch.1 will be followed by data of ch.3.
Output Format in case of Normal Measurement
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
V1 data
Terminator
Terminator
Terminator
Terminator
Terminator
A1 data
W1 data
Frequency
END
,
Display C
Note
•
When the frequency is set by either of the following methods, only one value is measured,
and that value will be output.
-
by panel keys: FUNCTION key on display C.
-
by communication command: by the “DC” or “EC” command.
After setting the measurement object of frequency, even changing the display C to something
different than VHz or AHz will not result in changing the object of measurement of frequency.
When selecting the output items yourself and you set a frequency item which is not object of
measurement, “999999.E
+
03” will be output.
•
The displayed values of V (voltage), A (current), W (active power), VA (apparent power), var
(reactive power), Vpk (voltage peak), and Apk (current peak) while the MAX hold function
(see section 4.8) is enabled will be the maximum values (MAX) that are held. The values
output via communications are also set to the maximum values (MAX) that are held.
10.4 Output Format