78
78
IM 54011-E
6.3
Scaling a voltage input range
The scaling settings for voltage inputs from channels A and B can be made using Thermo-
Collector (PC software).
■
What scaling is
The value Y converted from the input voltage level X according to the following equation is
displayed:
Y = aX + b.
Scaling setting means to set gain a and bias b.
For example, if you want to display:
• “0” when the input level is 0 V
• “200” when the input level is 1 V
then the gain and bias should be set as
•
a = 200
•
b = 0.
The resulting equation is
Y = 200X.
When
X is 0, Y is 0. When X is 1, Y is 200.
■
Setting procedure
See Also
For how to start Thermo-Collector on the PC, refer to Chapter 3, “Setting Up with the Personal Com-
puter” in the “Preparing for Measurement” part.
1.
From the File menu, choose Voltage Range Settings to open the Voltage Range Set-
tings dialog box.
[TIP]
The following shows the default settings.
Range
Name
Unit
Gain a
Bias b
1V
1V
V
1.0
0
100mV
100mV
mV
1.0
0
6 Changing Measuring Range
Input from sensor
1V
200
0
Displa
y at
TM20