65
65
IM 54011-E
Measuring
T
emperatures
When measuring temperatures with Thermo-Collector, it is necessary to use an
appropriate input channel and probe. By combining these items in different ways,
it is possible to measure the temperature of various objects. The temperature of
the surface and center of a solid, and various types of liquids can be measured.
For example, the “C” input channel (built-in sensor of the main unit) is used to
measure the ambient air temperature. To measure the temperature at the center
of a hamburger, connect a needle-type standard probe to the “A” measuring chan-
nel. To measure the surface temperature of a steel plate, connect a non-contact
thermo probe to the “D” input channel.
The following table shows the available probe types, their corresponding input
channels, and the positions of connection jacks.
Probe (option) type
Input channel
Position of connection jack
Thermo probe
Model name: 900 20
Thermo probe
Model name: 900 21
Thermo probe
Model name: 900 22
Thermo probe
Model name: 900 23
Thermo probe
Model name: 900 24
A, B
Top face (position marked “A”, “B” on the cover)
Thermo probe
Model name: 900 30
Thermo probe
Model name: 900 31
Thermo probe
Model name: 900 32
Thermo probe
Model name: 900 33
Beads TC
Model name: 2459 07
Non-contact thermo probe (Model name: 900 03)
D
Left side of the bottom face (position marked “D” on the cover)
Warning
Do not point the tip of a needle-type thermo probe towards somebody. The needle tip
is very sharp and can cause injury.
Caution
• Handle the needle-type thermo probe carefully so that its stylus is only put into the
intended measuring object.
• Do not use this instrument in a location where it may be splashed or dropped in
water, otherwise it may malfunction.
1
Input Channel and Probe
1 Input Channel and Probe