IM 54011-E
5 Confirm, Modify, Clear, or Delete the Data with the Main Unit
63
63
Preparing f
o
r
Measurement
5.10
Initializing the main unit
This section explains the procedure used to
initialize the main unit.
Note
Before initializing the main unit, it is rec-
ommended to save the setup data and
measurement data in the main unit to a
personal computer, etc.
When the main unit is initialized, all the
setup data and measurement data re-
corded in the main unit is lost. However,
the instrument name, chime setup, and
time setup data will not be erased.
See Also
For how to save the measurement data, refer to
Chapter 1, “Receiving the Data on the Personal
Computer” in the “Working with Measurement
Data” part.
Home screen
18:54:27
222
222
25.9
°
C
1.
Press the [FUNC], [0], and [1] keys, in
order.
The “Initial” screen appears on the LCD.
F01
I n i t i a l
2.
Press the [SET] key.
I n i t i a l ?
Ye s 1 / N o 0
3.
Press the [1] key.
The initialization process begins.
I n i t i a · ·
In two minutes, all the data set in the main unit is erased
and the “Initial” screen is restored,
Press the [ESC] key repeatedly to return to the Home
screen.
-End of procedure-
5.11
Confirming the current op-
erator name
This section explains the procedure used to
confirm the operator name currently set.
Home screen
18:54:27
222
222
25.9
°
C
1.
Press the [FUNC], [8], and [1] keys, in
order.
The “CurrData” screen appears on the LCD.
F81
C u r r D a t a
2.
Press the [SET] key.
The operator name currently set is displayed.
1
S m i t h
Press the [ESC] key repeatedly to return to the Home
screen.
-End of procedure-