4 Setting Up with the Main Unit
4.2
Setups required to use the
logging function
(setting the log name, input
channel, measurement in-
terval, and measurement
period)
Before using the logging function, it is nec-
essary to make logs settings.
■
Setup contents
See Also
For more information about each setup item,
refer to Section 3.2, “Setups required to use
the log (Setting the log name, input channel,
measurment interval, and measurement pe-
riod)”
•
Log name
•
Input channel
•
Measurement Interval
•
Measurement period
■
Operation procedure
Home screen
18:54:27
222
222
25.9
°
C
1
. Press the [FUNC], [2], and [3] keys, in
order.
The “Log Ins” screen appears on the LCD.
F23
L o g I n s
2
. Press the [SET] key.
A screen for selecting the log name appears.
Accept1
0
[TIP]
If no log has been created in the main unit,
[NewLog] is displayed on the screen.
3.
Press the [
▼
] key repeatedly until
[NewLog] appears on the screen.
[NewLog]
0
4
. Use the [CH] key to select the input
channel.
The channel indication below “INPUT” changes ac-
cordingly.
[NewLog]
0
[TIP]
Select “A” and “B” at the two-channel simul-
taneous measurement.
5
. Press the [SET] key.
A screen for entering a new log name appears.
[ N ew L o g ]
[TIP]
If a new log is created while [NewLog] is dis-
played, the created log is added to the end of
the list. If the [SET] key is pressed while
[NewLog] is not displayed, the created log is
inserted immediately before the selected log.
6.
Enter the log name.
Accept2
[ N ew L o g ]