7-1
IM 704610-01E
Signal Output
7
Chapter 7
Signal Output
7.1
DC Output of Jitter Ratio
<<For a functional description, see section 2.6>>
Procedure
MEDIA
FUNCTION
POLARITY
ARMING
INHIBIT
LOCAL
RECALL
ENTER
TRIG
IMPEDANCE EQUALIZER
SEC RANGE
KEY
LOCK
MEASURE
DISP OFF
JUDGE
GO
NO-GO
10%
20%
S
SEC RANGE
SCALE
L.MARK
R.MARK
GATE TIME
DELAY
BLOCK
REMOTE
STORE
UTILITY
JDG LEVEL
LEVEL
AUTO
COUPLE
EQ BOOST
MAN
CONV
50
Ω
1M
Ω
PW
E2T
D to C
BD x1
ns ms
s
%
V
%/
S
SHIFT
SHIFT
Display 1
Display 2
The phrase, “rotary knob & <>”
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.
You must connect a BNC cable from the jitter ratio DC output terminal (JITTER DC
OUT) on the rear panel of the instrument to a monitoring device before outputting
DC signals following the procedures in this section.
1.
Press SHIFT+ENTER (UTILITY). The Utility menu is shown on Display 2.
2.
Turn the rotary knob to display DC out in Display 2.
3.
Press >. DC mode, DCJitHi, DCJitLo, DC ave, Coeff•
α
, or Coeff•
β
is displayed.
Selecting the DC Output Mode
4.
Turn the rotary knob to select DC mode.
5.
Press > (or ENTER). Display 1 blinks.
6.
Turn the rotary knob to select Jitt or JudGE.
7.
Press ENTER.
If you select Jitt (jitter ratio DC output), a DC voltage corresponding to the jitter
ratio is output from the jitter ratio DC output terminal. If you select JudGE
(determination output), a DC voltage of 5 VDC is output from the jitter ratio DC
output terminal when the jitter ratio is equal to or below the determination level
and 0 VDC when the jitter ratio is above the determination level.
Setting the Jitter Ratio DC Output Range
• Setting the Upper Limit of the Jitter Ratio DC Output Range
4.
Turn the rotary knob to select DCJitHi.
5.
Press > (or ENTER). Display 1 blinks.
6.
Use rotary knob & < > to set the upper limit.
7.
Press ENTER.
• Setting the Lower Limit of the Jitter Ratio DC Output Range
4.
Turn the rotary knob to select DCJitLo.
5.
Press > (or ENTER). Display 1 blinks.
6.
Use rotary knob & < > to set the lower limit.
7.
Press ENTER.