10-26
IM 704610-01E
:MEMory:SEND?
Function
Executes the transmission of the measured data
specified by “MEMory:DATaselect.”
Syntax
:MEMory:SEND?
Example
:MEMORY:SEND?
-> #800000016abcdabceabcfabcg
Description • When the transmission format is ASCII,
response data of only the number of data
points are output in <NR3> format delimited
with commas. For BINARY, the output is
simply block data.
• Conversion of binary data to real numbers is
as follows.
• Frequency (FREQuency):
The frequency is stored using a 4-byte
unsigned integer.
• Measured values (MEASuredata):
4-byte signed integer. The measured value
is derived by multiplying by 97.65625 s.
The frequency outside of the
HHISt:Marker:LEFT and
HHISt:Marker:RIGHt range is 0.
This command can only be used after the
single measurement command (sstart).
This command cannot be used after the
start or stop command.
10.2 Commands
:MEMory:STARt
Function
Sets the start position of transmitted data or
queries the current setting.
Syntax
:MEMory:STARt {<time>}
:MEMory:STARt?
<time> = 0 to 80 ns (when the measurement
function is set to D-to-C)
<time> = 0 to 999.9 ns (when the measurement
function is set to PulseWidth)
Example
:MEMORY:START 1ns
:MEMORY:START?
-> :MEMORY:START 1.0E-9
10.2.11 RECall Group
:RECall
<Space>
<NRf>
:RECall
Function
Recalls the setup information.
Syntax
:RECall {<NRf>}
<NRf> = 0 to 6
Example
:RECALL 0