11-3
IM 704610-01E
Other Functions
11
11.2 Initializing Settings
Explanation
The following setup information can be initialized to the factory default settings.
Item
Factory Default Settings
Measurement Conditions
Measured media
BD (fixed)
Measurement function
D to C
Lower limit of the pulse width measurement (L.MARK)
22.72 ns
Upper limit of the pulse width measurement (R.MARK)
37.88 ns
Polarity of the data signal
Gate time
10 ms
Arming type
Auto arming (internal arming)
Slope of the external arming signal
Unselected
Arming delay setting
0.0 ms
Block sampling setting values
1
ON or OFF condition of inhibit
OFF
Polarity of the inhibit signal
Unselected
ON or OFF condition of the equalizer
OFF
Boost amount of the equalizer
CONV: 5.8 dB
Trigger mode type
AUTO (auto mode)
Slice level
0
Offset level
0
Input Impedance
1 M
Ω
Input coupling
DC
Display
Meter scale type
10%
Time range of meter scale
0.5 ns
Numerical value display parameters
σ
/T (jitter ratio)
Numerical/alphabetical character display ON or OFF
ON
Jitter ratio determination level (GO/NO-GO indicator)
10.00%
Store/Recall memory number 0
UTILITY-DC out
DC output mode
DC output of jitter ratio
Upper limit of the jitter ratio output range
25.00%
Lower limit of the jitter ratio output range
0.00%
Average coefficient of the DC output filter
1
Jitter ratio correction coefficient
α
1.0000
Jitter ratio correction coefficient
β
0.000%
UTILITY-Level
Voltage level measurement ON or OFF
OFF
Voltage level DC output mode
Voltage level DC output
Upper limit of the voltage level DC output range
5.000 V
Lower limit of the voltage level DC output range
0.000 V
Upper limit of the voltage level determination range
1.000 V
Lower limit of the voltage level determination range
0.000 V
Average coefficient for DC output of the voltage level
1
UTILITY-Meas
PLL hold ON/OFF
OFF
Turning D-Ccalc H-ON or H-OFF
H-OFF
(there are no D-Ccalc settings on products with suffix code -BDS.)
AGC ON or OFF
ON
ON or OFF condition of the numerical display
OFF
UTILITY-Display
Turning display of
σ
/T ON or OFF
ON
Turning display of
σ
ON or OFF
ON
Turning display of AVE ON or OFF
ON
Turning display of T ON or OFF
ON
Turning display of SnU ON or OFF
OFF
Turning display of SnL ON or OFF
OFF
Turning display of EQ ON or OFF
ON
Turning display of Dly ON or OFF
ON