Explanation of Functions
2
Inhibit <<For procedures, see section 4.8>>
You can inhibit measurements by applying an external signal to the inhibit signal input
terminal. This is possible even while the gate is open or during measurement after
arming activation. You can also select which polarity of the signal, positive or negative, is
used to inhibit measurements. The relation between the inhibit signal, gating, and
arming for pulse width jitter measurement is indicated below.
Relation between the Inhibit Signal and Gating
S1
S2
S3
S4
S5
Data signal
Not measured
Gate
Inhibit
Inhibits measurement
Gate open
Starts measurement
Resumes measurement
Relation between the Inhibit Signal, Gating, and External Arming
S1
S2
S3
S4
S5
Data signal
Not measured
External arming
Gate
Inhibit
Inhibits measurement
Gate open
Starts measurement
Resumes measurement
Relation between the Inhibit Signal, Gating, External Arming, and Arming Delay
S1
S2
S3
S4
S5
Data signal
Arming delay
Not measured
External arming
Gate
Inhibit
Inhibits measurement
Gate open
Starts
measurement
Resumes
measurement
2.4 Measurement Conditions