10-25
IM 704610-01E
Communication Command
10
10.2 Commands
10.2.10 MEMory Group
The commands in this group deal with the external transmission of the measured data.
There are no front panel keys that correspond to this group.
:MEMor y
:
BYTeorder
<Space>
LSBFirst
MSBFirst
?
DATaselect
<Space>
MEASuredata
FREQuency
?
END
<Space>
<Time>
?
FORMat
<Space>
ASCii
BINary
?
STARt
<Space>
?
SEND
?
?
<Time>
;
CLEar
:MEMory?
Function
Queries all settings related the external
transmission of the measured data.
Syntax
:MEMory?
Example
:MEMORY?
-> :MEMORY:BYTEORDER LSBFIRST
;DATASELECT FREQUENCY;END 15.1E-09
;FORMAT ASCII;START 1.0E-09
:MEMory:BYTeorder
Function
Sets the transmission order of binary data or
queries the current setting.
Syntax
:MEMory:BYTeorder {LSBFirst|
MSBFirst}
:MEMory:BYTeorder?
Example
:MEMORY:BYTEORDER LSBFIRST
:MEMORY:BYTEORDER?
-> :MEMORY:BYTEORDER LSBFIRST
:MEMory:CLEar
Function
Clears the measured data.
Syntax
:MEMory:CLEar
Example
:MEMORY:CLEAR
:MEMory:DATaselect
Function
Sets the data to be transmitted or queries the
current setting.
Syntax
:MEMory:DATaselect {MEASuredata|
FREQuency}
:MEMory:DATaselect?
Example
:MEMORY:DATASELECT FREQUENCY
:MEMORY:DATASELECT?
-> :MEMORY:DATASELECT FREQUENCY
:MEMory:END
Function
Sets the end position of transmitted data or
queries the current setting.
Syntax
:MEMory:END {<time>}
:MEMory:END?
<time> = 0 to 80 ns (when the measurement
function is set to D-to-C)
<time> = 0 to 999.9 ns (when the measurement
function is set to PulseWidth)
Example
:MEMORY:END 1 ns
:MEORY:END?
-> :MEMORY:END 1.0E-09
:MEMory:FORMat
Function
Sets the format of the data to be transmitted or
queries the current setting.
Syntax
:MEMory:FORMat {ASCii|BINary}
:MEMory:FORMat?
Example
:MEMORY:FORMAT ASCII
:MEMORY:FORMAT?
-> :MEMORY:FORMAT ASCII