4-7
IM 704610-01E
Measurement Condition Settings
4
4.5
Setting the Gate Time
<<For a functional description, see section 2.4>>
Procedure
MEDIA
FUNCTION
POLARITY
ARMING
INHIBIT
LOCAL
RECALL
ENTER
TRIG
IMPEDANCE EQUALIZER
SEC RANGE
KEY
LOCK
MEASURE
DISP OFF
JUDGE
GO
NO-GO
10%
20%
S
SEC RANGE
SCALE
L.MARK
R.MARK
GATE TIME
DELAY
BLOCK
REMOTE
STORE
UTILITY
JDG LEVEL
LEVEL
AUTO
COUPLE
EQ BOOST
MAN
CONV
50
Ω
1M
Ω
PW
E2T
D to C
BD x1
ns ms
s
%
V
%/
S
SHIFT
SHIFT
Display 1
Display 2
The phrase, “rotary knob & <>”
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.
1.
Press SHIFT+FUNCTION (GATE TIME). GateTime appears in Display 2, and
the setting value appears in Display 1.
2.
Use rotary knob & < > to set the gate time.
Explanation
You can set the time (gate time) during which the measured values of pulse width and
phase difference are stored in the acquisition memory.
Gate Time
The duration can be set within the following range.
Setting range
When D-to-C high speed calculation (see section 4.9) is OFF:
1 to 1000 ms (in 1 ms steps)
When D-to-C high speed calculation is ON:
2 to 1000 ms (in 2 ms steps)
Note
Loading the Measured Values When D-to-C High Speed Calculation is OFF
•
When the gate time is set between 100 ms and 1000 ms, the measured values are acquired
in units of 100 ms. The measured values are acquired so that the sum of the gate times in
units of 100 ms add up to the specified gate time, and those values are used to determine the
jitter (statistical value). The calculated value is displayed or output as a DC level signal (see
section 7.1). Meter indication, numerical display, and DC output are updated as shown below.
When measurement is performed only once using the SSTart communication command, if
external arming (see section 4.6) is selected, the acquisition is performed all at once without
being divided up by the specified gate time in 100 ms units.
RF
signal
100 ms
100 ms
100 ms
100 ms
100 ms
100 ms
50 ms
50 ms
Meter indication, DC output
Example in which the gate time is set to 250 ms
50 ms
RF
signal
100 ms
100 ms
100 ms
100 ms
100 ms
100 ms
50 ms
50 ms
Numerical display function
50 ms