5-8
IM DL350-03EN
System Errors (900 to 999)
Code
Message
Chapter or
Section
901
Failed to backup setup data.
Will initialize.
Backup battery may be low.
–
903
The USB device's power consumption exceeded the capacity of the USB hub.
–
904
Lower the sample rate or reduce the number of measuring channels.
Chapter 2
1
905
Lower the sample rate or reduce the number of measuring channels.
Chapter 2
1
906
Fan stopped.
Maintenance service is required.
It will shutdown.
–
907
Internal temperature is too high.
Maintenance service is required.
It will shutdown automatically.
–
909
SD Card is full.
6.10
1
910
Key protect is enabled.
To release the protection, press the PROTECT key or enter the password.
20.5
1
912
Fan for CPU stopped.
Maintenance service is required.
It will shutdown automatically.
–
913
LCD BackLight Failure.
Maintenance service is required.
–
915
It installed the module which cannot support by this machine.
–
917
Hardware configuration error occurred.
Restart this machine.
If it occurred again, maintenance service is required.
2.9
918
Error occurred while ImageFile process.
–
919
Key operate not available while system error occurred
–
920
Firmware overwriting error occurred.
–
922
GPS module does not work.
2.22
923
Setting was changed.
Will initialize.
–
930
Slot 1 EEPROM error.
–
931
Slot 2 EEPROM error.
–
938
Internal flash write error.
–
939
Internal flash verify error.
–
940
Because abnormal temperature was detected, then it will shut down.
–
941
Because an unspected battery was detected, then it will shut down.
2.5
942
Because USB overcurrent was detected, the USB power was turned off.
Please disconnect the connected USB device and restart it.
2.9
943
Because GPS overcurrent was detected, the GPS power was turned off.
Please disconnect the GPS unit and restart it.
2.22
944
Because module insertion/removal was detected, acquisition was stopped.
Please restart it.
2.9
945
Low Battery.
Please connect to another power supply immediately.
2.6, 2.7
946
Because a battery residual quantity disappeared, It will shutdown.
2.9
947
Battery voltage is low.
Please connect to another power supply immediately.
2.6, 2.7
1 User’s Manual, IM DL350-02EN
Note
If servicing is required, first see if initializing the instrument fixes the problem.
5.2 Messages and Corrective Actions