Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
App
Index
Code
Message
Chapter or
Section
850
Cannot be set when the trace is set to a frequency module.
–
852
Cannot set Math to OFF while FFT Window ON.
10.1
1
853
Cannot select this trace because it already selected.
–
854
Because a record length is too long, it can't be set up by the present number of indication channels.
Chapter 2,
1
3.1
1
855
Cannot change to such Record length while running.
Set the trigger mode to Auto and decrease T/Div to less than 100 msec/div to enable roll mode, or set
the trigger mode to Single or On Start.
–
856
Cannot Display setting to On.
This CH didn't acquisition to memory.
Chapter 2
2
858
Cannot set while action mode is on.
3.1
1
868
Cannot be specified because characters in the JIS level-2 kanji set are included.
Create the file on the local drive, and then copy it to the network drive.
–
869
Cannot set while Go-Nogo mode.
Turn Off Go-Nogo mode first.
Chapter 13
1
870
All sub-channel inputs are off.
At least one inputs should on.
2.3,
1
2.5
1
871
No effective channel for Math Setup.
–
875
Cannot change or START when accumulate is ON.
Turn OFF accumulate first.
5.3
1
879
Cannot set GoNogo mode while Math or FFT Window is On.
Chapter 9,
1
chapter 10
1
881
Cannot set for CH which ValueType is Float while running.
–
883
Cannot set Input to ON with limit of memory capacity.
2.10,
1
2.11,
1
2.12
1
885
Zoom rate is limited while installed a module which has sub-channels.
–
890
LIN Monitor does not support Float data.
2.11
1
892
Cannot set Math operation while FFT window is open.
10.1
1
893
Cannot set that item.
It cannot be measured for Logic module.
8.1
1
894
Cannot set DC Offset Cancel to ON due to following case.
- Coupling is not DC.
- Not execute DC Offset Cancel.
2.2,
1
3.6
895
The password entered the first time is different from the password entered the second time.
Reenter the password for the second time.
–
896
It is not possible to set the subchannel sample interval to be 1 second or more.
(sample rate to be 1S/s or less.)
3.1,
1
3.2
1
897
This function cannot be used when
the recorder mode is set.
–
898
Cannot change when Internal Clock is active.
–
1 User’s Manual, IM DL350-02EN
2 Features Guide, IM DL350-01EN
5.2 Messages and Corrective Actions