5-17
5. Parameter Setting
Code
Parameter spewcific functions
01
02
03
04
05
06-09
Display
*SC.RES
*4-ELEC
*0.10xC
*AIR
*START
*"WAIT"
*END
*POL.CK
Function
Select main parameter
Select 2/4-EL system
Set celconstant
Zero calibration
Polarisation check
Function detail
Conductivity
Resistivity
2-Electrode measurement system
4-Electrode measurement system
Press NO to step through choice of
multiplying factors on the second display.
0.10xC
1.00xC
10.0xC
100.xC
0.01xC
Press YES to select a factor
Use >, , ENT keys to adjust MAN digits
Zero calibration with dry cell connected
Press YES to confirm selection
Press YES to start, after briefly displaying
"WAIT", *END will be displayed
Press YES to return to commissioning mode
Polarisation check off
polarisation check on
Not used
X
0
1
0
1
0
1
Y Z Default values
0
Cond.
0
2-El.
0.100
cm-1
0.10xC
1.000
1
On