3. Installation and Wiring
3-3-3. AC power
Connect terminal 1 to the phase line of the AC power and terminal 2 to the zero line.
Terminal 3 is for the power ground. This is separated from input ground by a galvanic
isolation.
3-3-4. Grounding the housing
To protect the instrument against interference, the housing should be connected to
ground by a large area conductor. This cable can be fixed to the rear of the housing
using a braided wire cable. See figure 3-8.
3-3-5. Switching on the instrument
After all connections are made and checked, the power can be switched on from the
power supply. Make sure the LCD display comes on. All segments will illuminate, and
then the instrument will momentarily display its unique serial number. After a brief
interval, the display will change to the measured value. If errors are displayed or a valid
measured value is not shown, consult the troubleshooting section (Chapter 8) before
calling Yokogawa.
Figure 3-8. Grounding the housing