IM 12D08N04-01E
4-1
4. Operation; Display Function and Setting
4.
Operation; Display Functions and
Setting
4-1. Operator interface
This section provides an overview of the operation of the EXA operator interface. The
basic procedures for obtaining access to the three levels of operation are described
briefly. For a step-by-step guide to data entry, refer to the relevant section of this
instruction manual. Figure 4-1 shows the EXA operator interface.
LEVEL 1: Maintenance
These functions are accessible by pushbutton through a flexible front cover window. The
functions make up the normal day-to-day operations that an operator may be required to
complete. Adjustment of the display and routine calibration are among the features
accessible in this way. (See table 4-1).
LEVEL 2: Commissioning
A second menu is exposed when the EXA front cover is removed and the display board
is revealed. Users gain access to this menu by pressing the button marked * in the lower
right of the display board. This menu is used to set such values as the output ranges and
hold and wash features. It also gives access to the service menu. (See table 4-1).
LEVEL 3: Service
For more advanced configuration selections, press the button marked * , then press
“NO” repeatably until you reach SERVICE. Now push the “Yes” button. Selecting and
entering “Service Code” numbers in the commissioning menu provide access to the
more advanced functions. An explanation of the Service Codes is listed in chapter 5 and
an overview table is shown in chapter 10.
Table 4-1. Operations overview
Routine
Function
Chapter
Maintenance
SETPOINTS Adjust alarm setpoints (when activated)
5
CALIB
Calibration with a standard solution or a sample
6
DISPLAY
Read auxiliary data or set message display
4
HOLD
Switch hold on/off (when activated)
5
Commissioning
SETPOINTS Adjust alarm setpoints
5
RANGE
Adjust the output range
5
SET HOLD
Activate the output range
5
TEMP
Select method of temperature compensation
5
Service
SERVICE
Fine tune the specialised functions of the converter 5
(Access to coded
entries from the
commissioning level)
NOTE
All three levels may be separately protected by a password. See Service Code 52 in
chapter 5 Service Code table for details on setting passwords.