IM 12D08N04-01E
2-1
2. SC402G Specifications
2.
SC402G Specifications
2-1. General
A.
Input specifications
: Two or four electrodes measurement with square wave excitation, using cell constants
from 0.008 to 50.0 cm
-1
B.
Detection method
: Frequency, read-pulse position and reference voltage are dynamically optimized.
C.
Input ranges
Conductivity
Minimum
: 0
µ
S/cm
Maximum
: 200 mS
3
(Cell constant) (overrange 1999 mS/cm).
Resistivity
Minimum
: 0.005 k
Ω
/ (Cell constant)
Maximum
: 999 M
Ω
3
cm
Temperature
Pt1000
: -20 to 250
°
C (0 to 500
°
F)
Pt100/Ni100
: -20 to 200
°
C (0 to 400
°
F)
8.55k
Ω
NTC
: -10 to 120
°
C (10 to 250
°
F)
PB36 NTC
: -20 to 120
°
C (0 to 250
°
F)
D.
Span
Conductivity
:min. 0.010
µ
S/cm,
max. 1999 mS/cm. (max 90% zero suppression)
Resistivity
:min. 0.001 k
Ω
·cm
max. 999 M
Ω
·cm. (max 90% zero suppression)
Temperature
: Dependent on temp. sensor type:
Sensor type
min.
max.
Pt1000
25
°
C (50
°
F)
250
°
C (500
°
F)
Pt100, Ni100
25
°
C (50
°
F)
200
°
C (400
°
F)
PB36NTC, 8.55k
Ω
NTC 25
°
C (50
°
F)
100
°
C (200
°
F)
The instrument is user programmable for linear or non-linear conductiv-
ity ranges.
E.
Transmission signals
: Two isolated outputs of 0/4-20 mA DC with common negative.
Maximum load 600
Ω
.
Auxiliary output can be chosen from conductivity, resistivity or temperature
Burn up (22 mA) or Burn down (0/3.5 mA) to signal failure.
F.
Temperature compensation
: Automatic, for temperature ranges mentioned under C (inputs).
Reference temp : programmable from 0 - 100
°
C or 30 - 210
°
F (default 25
°
C).