IM 12D08N04-01E
6-2
6. Calibration
6-2. Calibration procedure
S / c m
ENT
S / c m
ENT
S / c m
ENT
S / c m
NO
YES
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
MEASURE
DISPLAY
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
CALIBRATE
SETPOINTS
S / c m
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
S / c m
NO
YES
S / c m
NO
YES
YES
MODE
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES to return to the measuring mode.
Note:
If calibration is out of limits (+/- 20% of Nom. Cell Constant), " Error 3" will appear in
the display.
To continue press "YES" or "NO".
" CAL. Cell Constant." will be shown in the display. Press "YES" to calibrate the cell
constant. (New cell constant is calculated and updated).
Press NO to stop calibration of the cell constant. (No update of cell constant.)