6-3
6. Calibration
6-3. Calibration with HOLD active
S / c m
ENT
S / c m
ENT
S / c m
ENT
S / c m
NO
YES
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
MEASURE
DISPLAY
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
CALIBRATE
SETPOINTS
S / c m
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
S / c m
NO
YES
S / c m
NO
YES
HOLD
YES
MODE
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
HOLD
HOLD
HOLD
HOLD
HOLD
S / c m
NO
YES
HOLD
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.