IM 12D08N04-01E
8-2
8. Troubleshooting
The following error message table gives a list of possible problems that can be indicated
by the EXA.
Table 8-1. Error Codes
Possible cause
Sensor surface fouled
Conductivity too high
Incorrect field calibration of TC
Calibrated value differs more than
+/- 20 % of nominal value programmed
in code 03.
Wrong data entered in 5x5 matrix
Incorrect wiring
Internal leakage of sensor
Defective cable
Dry sensor
Incorrect wiring
Defective cable
Process temperature too high or too low
Wrong sensor programmed
Incorrect wiring
Process temperature too high or too low
Wrong sensor programmed
Incorrect wiring
Too high zero due to cable capacitance
Fault in electronics
Poor water quality
Cable resistance too high
Corroded contacts
Wrong sensor programmed
Incorrect configuration by user
Wrong data programmed
Incorrect configuration by user
Fault in electronics
Very severe interference
Software problem
Process control not effective .
within set time
Code
E1
E2
E3
E4
E5
E6
E7
E8
E9
E10
E13
E15
E17
E18
E19
E20
E21
E22
Error description
Polarisation detected on cell
Temperature coefficient out of limits
(exceeds 0~3.5%/
°
C range)
Calibration out of limits
Matrix compensation error
Conductivity too high or resistivity too low
(Limits set in service code 54)
Conductivity too low or resistivity too high
(Limits set in service code 54)
Temperature sensor open
(Pt1000 : T > 250
°
C or 500
°
F)
(Pt100/Ni100 : T > 200
°
C or 400
°
F)
(8.55k
Ω
NTC : T < -10
°
C or 10
°
F)
(PB36 NTC : T < -20
°
C or 0
°
F)
Temperature sensor shorted
(Pt1000/Pt100/Ni100 : T < -20
°
C or 0
°
F)
(8.55k
Ω
/PB36 NTC : T > 120
°
C or 250
°
F)
Air set impossible
EEPROM write failure
USP limit exceeded
Cable resistance influence to temperature
e/- 15
°
C
Output span too small
Table values make no sense
Programmed values outside acceptable limits
All programmed data lost
Checksum error
Alarm activation time exceeded
Suggested remedy
Clean sensor and calibrate
Replace sensor
Re-adjust
Set calculated TC
Check for correct sensor
Check for correct unit (
µ
S/cm,
mS/cm, k
Ω
.cm or M
Ω
.cm)
Repeat calibration
Re-program
Check wiring (3-6)
Replace sensor
Replace cable
Immerse sensor
Check wiring (3-6)
Replace cable
Check process
Check model code sensor
Check connections and cable
Check process
Check model code sensor
Check connections and cable
Replace cable
Try again, if unsuccessful
contact Yokogawa
Check process
Check cable
Clean and reterminate
Reprogram
Reprogram
Reprogram
Reprogram
Contact Yokogawa
Contact Yokogawa
Check control equipment
Adjust value in code 47