3-8
IM PX8000-02EN
3.8 Triggering on an A -> B(N) Trigger (Enhanced)
This section explains the following settings for triggering on an A -> B(N) trigger.
• Trigger source
• State condition
• State condition achievement condition
• Number of times state condition B must be met
• Trigger condition
►
Features Guide: “A -> B(N) Trigger (Enhanced)”
ENHANCED_A->B(N) Trigger Menu
Press
SIMPLE/ENHANCED
and then the
Setting
soft key to select Enhanced. The following menu
appears.
Set Setting to Enhanced.
Set the trigger hold off. ► section 3.3
Set Type to A -> B(N).
Set the state and trigger conditions.
Setting the State and Trigger Conditions (Set Pattern)
Press the
Set Pattern
soft key to display the following menu.
Set the state condition (H, L, or X (do not use as a trigger source)).
Set the trigger level.
Set the hysteresis ( , , ).
Set the number of times state condition B
must be met.
Set the state condition achievement conditions
(Enter, Exit).
3.8 Set anchor at “enhanced
trigger.”