2-18
IM PX8000-02EN
AUX Menu
1.
Press
U2(CH3)
,
U3(CH5)
, or
U4(CH7)
to select a channel on an installed AUX module. The
following menu appears.
Turns the waveform display on
and off
Set the display label.
Set the zoom method to DIV.
Set the zoom magnification.
Set the offset.
Turns the auto range on and off*
Turns the waveform display on
and off
Set the display label.
Set the zoom method to SPAN.
Set the upper and lower limits of
the display range.
Turns the auto range on and off*
When the zoom method is
set to SPAN
When the zoom method is
set to DIV
Displays the second
page of the menu
Configure rotating
speed measurement.
Configure rotating
speed measurement.
Displays the second
page of the menu
* You can turn auto range on and off when the rotating-speed input signal type is set to Analog. If the type
is set to Pulse, auto range is fixed to OFF.
Note
The U key whose display setting is ON illuminates. If the U key is not illuminated, you can press it to turn on
the waveform display and the key. If the U key is illuminated, you can press it to turn off the waveform display
and the key.
2.
Press the
Next 1/2
soft key to display the 2/2 menu.
Set the input coupling (AC, DC,
GND).
Set the bandwidth limit (10kHz,
20kHz, 40kHz, 80kHz, 160kHz,
320kHz, 640kHz, 1.28MHz, 2MHz,
Full).
Set the upper and lower pulse
reference levels.
When the rotating-speed input signal
type is Analog
When the rotating-speed input signal
type is Pulse
Displays the first page
of the menu
Displays the first page
of the menu
Set the bandwidth limit (10kHz,
20kHz, 40kHz, 80kHz, 160kHz,
320kHz, 640kHz, 1.28MHz, 2MHz,
Full).
2.5 Configuring Rotating Speed Measurements