<9. In-Process Operation>
9-3
IM 01S01C01-01EN
Table 9.2
Field Diagnostic Alert
Indication of FD_*_ACTIVE
Indication of FD_RECOMMEN_ACT
Solution
Electronics failure
Repair electronics
Replace electrical parts e.g. amplifier.
Or contact sales office or service
center.
Sensor/Actuator failure
Repair Sensor/Actuator
Replace mechanics e.g. sensor or
actuator. Or contact sales office or
service center.
Potential failure
Investigate failure
Perform reconfiguration, cleaning,
wiring/connector or electrical board
check.
If alarm still persists, contact sales
office or service center.
Backup function in operation
Repair primary side
Repair primary sensor before backup
sensor fails.
Firmware update error
Retry updating firmware
Retry firmware update. Check cause of
the failure if alarm persists.
Communication configuration error
Configure communication correctly
Correct configuration of
communication.
Non-operating-state
Wait for a while
Wait for a while. Check cause of the
failure if alarm persists.
Calibration warning
Check calibration
Investigate cause of failure and
recalibrate device.
Device configuration error
Configure device correctly
Correct configuration relating to sensor
or actuator.
Function restricted
Confirm the state
Check if this is right state.
Simulation mode
Confirm the state
Check if this is right state.
Manual mode
Confirm the state
Check if this is right state.
Function Block notice
Check Function Block status
Check conditions of function
blocks. In order to avoid alarm from
unused function blocks, configure
RESOURCE2.FD_EXTENDC_MAP_n
(n: 1 to 3) parameter.
Sensor/Actuator out of range
Check specification
Check specification of sensor and
actuator. Or process conditions may be
temporarily non-conforming.
Out of operating limit
Check environment
Check environment specification
of sensor and actuator. Or process
environment may be temporarily non-
conforming.
Temporal decrease of value quality
Check process or peripherals
Check process and peripherals
conditions.
Deterioration estimated by Time Based
Maintenance
Check deterioration
Check if maintenance is required.
Deterioration estimated by Condition
Based Maintenance
Check deterioration
Check if maintenance is required.
Optional function configuration error
Check optional configuration
Check configuration of optional
functions.
Alarm related information
Confirm information
Check the alarm related information.
Process alarm
Check process
Check process conditions.
Содержание FVX110
Страница 1: ...User s Manual FVX110 Fieldbus Segment Indicator IM 01S01C01 01EN IM 01S01C01 01EN 4th Edition ...
Страница 15: ... 2 Handling Cautions 2 4 IM 01S01C01 01EN ...
Страница 16: ... 2 Handling Cautions 2 5 IM 01S01C01 01EN ...
Страница 17: ... 2 Handling Cautions 2 6 IM 01S01C01 01EN ...
Страница 18: ... 2 Handling Cautions 2 7 IM 01S01C01 01EN ...
Страница 19: ... 2 Handling Cautions 2 8 IM 01S01C01 01EN ...
Страница 20: ... 2 Handling Cautions 2 9 IM 01S01C01 01EN ...