98
IM FG410-01EN
4. BASIC OPERATION
r) To substitute center value to marker value, or
substitute marker value to center value
When the [Ctr Mkr] soft key on the 2nd page of the setting screen is pressed, the center value
is substituted to the marker value. When the [Mkr Ctr] soft key is pressed, the marker value is
substituted to the center value.
When these soft keys are not displayed on the 2nd page of the setting screen, press the
[▼1/2] soft key on the right end to switch the soft key sets.
s) To control sweep start, stop, and pause with external logic signal
When [ExtCtrl] (external control) on the 3rd page of the setting screen is set to [Enable], sweep
manipulation through TTL level logic input to the multi-I/O connector on the rear panel can be done.
This applies in common to CH1 and CH2.
To prevent malfunction due to external noise, it is recommended to set [Disable] (disable) when not
using control with an external signal.
For the pin number allocation,
p.20 .
The following manipulations can be done.
Sweep start
If sweep operation is currently stopped, the sweep operation starts from the beginning upon input
of a falling edge.
During sweep execution, the sweep operation is resumed from the beginning upon input of a
falling edge.
During single-shot sweep and gated single-shot sweep, the sweep operation starts from the
beginning if a trigger other than this signal is received. ORing with the set trigger source is done.
Sweep stop
The sweep operation stops upon input of a falling edge during sweep execution, and the state
becomes the sweep start value output state.
However, during single sweep and gated single sweep, the sweep operation starts again from the
beginning upon receipt of a new trigger.
Sweep hold/resume
The sweep operation is held upon input of a falling edge during sweep execution. The sweep
operation resumes from where it was held upon input of a rising edge during hold.
However, during single-shot sweep and gated single-shot sweep, the sweep operation starts again
from the beginning upon receipt of a new trigger during hold.
Содержание FG420
Страница 10: ...VIII IM FG410 01EN...
Страница 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Страница 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Страница 167: ...153 IM FG410 01EN RACK MOUNTING DIMENSIONS Inch rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Страница 169: ...155 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Страница 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Страница 173: ...159 IM FG410 01JA...
Страница 178: ......
Страница 179: ......