17
IM FG410-01JA
3.2.4
External
trigger input (
TRIG IN)
CH2
TRIG IN
TRIG IN
CH1
TRIG IN
FG420
FG410
TRIG IN may be used as external trigger input in the following cases. The polarity setting can be
changed.
• Single-shot sweep start trigger
• Gated single-shot sweep start trigger
• Trigger burst oscillation start trigger
• Gate of gated oscillation
• Triggered gate oscillation trigger
• Sequence oscillation start trigger
“6.2 Basics” in the Application Instruction Manual
• TRIG IN can also be used as the external modulation input for FSK and PSK
Input Characteristics
Input voltage
TTL level (low: 0.8 V or lower; high: 2.6 V or higher)
Maximum allowed input
-0.
5 V to +5.5 V
Input impedance
10 k
Ω
, pull up to +3.3 V
Signal GND
Same potential as housing
3.2.5 External 10 MHz frequency reference input (10 MHz REF IN)
FG420
FG410
10MHz REF IN
[Insulated from housing]
10MHz REF IN
[Insulated from housing]
10 MHz REF IN can be used for the following purposes.
3.2 I/O Terminals
Содержание FG420
Страница 10: ...VIII IM FG410 01EN...
Страница 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Страница 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Страница 167: ...153 IM FG410 01EN RACK MOUNTING DIMENSIONS Inch rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Страница 169: ...155 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Страница 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Страница 173: ...159 IM FG410 01JA...
Страница 178: ......
Страница 179: ......