112
IM FG410-01EN
4. BASIC OPERATION
f) To start auto burst → Automatically starts
When, in the auto burst mode, the mode is set to the burst oscillation mode, burst starts automatically.
However, if the burst setting is improper, burst oscillation will not start. ([Conflict!] is displayed in
the top right part of the screen.) When the [?] soft key appearing at the left end is pressed, a message
about the inappropriate setting is displayed. Once the setting has been changed to an appropriate
setting, burst oscillation starts.
g)
Stopping auto burst →Not possible
Oscillation cannot be stopped in the auto burst mode as is.
To stop oscillation, set [BrstMode] (burst mode) on the 2nd page of the setting screen to a setting
other than [Auto] (auto), and then set so that no trigger or gate signal is input.
To set continuous oscillation, change [Mode] (oscillation mode) to [Continuous].
h) To output burst sync signal: Sync output setting
Set burst sync signal with [SyncOut] (sync output) on the 2nd page of the setting screen. One of the
following two settings can be selected.
• Signal synchronized with the reference phase of the waveform [Sync]
• Signal synchronized with burst oscillation [BrstSync]
When [Sync] is selected
A TTL level signal that rises at the reference phase of the waveform is output from the sync/sub-
output terminal.
When [BrstSync] is selected
A TTL level signal synchronized with the burst oscillation is output from the sync/sub-output
terminal. As shown in the following figure, this signal is low during oscillation and high during
oscillation stop.
When the signal during burst is observed with an oscilloscope, it can be used as the trigger signal
for the oscilloscope.
Oscillation
Stop
Burst sync signal
Output signal
Oscillation
Stop
i) Usage of stop level
The level when oscillation is stopped is normally set with the oscillation start/stop phase, but it can
also be set independently from that with a proportion of the amplitude full scale.
Set [StpLvl] (stop level) on the 2nd page of the setting screen to [On] and set the level with a % value.
Содержание FG420
Страница 10: ...VIII IM FG410 01EN...
Страница 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Страница 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Страница 167: ...153 IM FG410 01EN RACK MOUNTING DIMENSIONS Inch rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Страница 169: ...155 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Страница 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Страница 173: ...159 IM FG410 01JA...
Страница 178: ......
Страница 179: ......